3510型原子吸收分光光度计系统 u 经济的主机
u 高灵敏度高稳定性的石墨炉系统
u 优秀的光学系统,确保测量数据准确可靠
u 火焰法测试安全可靠
u 高效率的分析工作站
产品用途
3510主机具有内置微机系统,所有工作条件通过键盘设定,仪器能自动寻找元素峰值能量,北京校正时空新阴极灯与氘灯的能量自动平衡,并能处理数据和图谱,真正体现了原子吸收分析系统快速、准确、灵敏的特点
产品参数
波长范围:190.0nm-860.0nm
波长准确度:全波段≤±0.5nm
波长重复性:≤ 0.3nm
闪耀波长:250nm
光谱带宽:0.1nm.0.2nm,0.7nm,1.4nm(开关选择)
分辨率:光谱带宽0.2nm,能分开锰双线,即279.5nm和279.8nm,且二谱线间的峰谷能量值小于40%
读书方式:连续、保持、峰高、峰面积
背景扣除:在背景信号为1ABS时具有30倍以上的背景扣除能力
吸光度范围:0.000A-2.000A
稳定性:在正常情况下,仪器和铜灯预热半小时后,在不点火状态下对铜谱线324.7nm测定基线稳定性。30分钟漂移量≤0.004吸光度
典型元素测定:火焰法,5mg/L的Cu吸光度大于0.800A
墨炉法灵敏度:Cu1×10-11g(特征量)
精密度:(相对标准偏差): ≤1%; ≤3%(Cu石墨炉测试)
特别申明:⑴本网站所列价目,如厂方有变动,请以厂家新价格为准。
⑵批发优惠幅度因各厂情况不同,请来电咨询。
Analytical Instruments Co., Ltd. Shanghai Jinpeng excellent chemical and biological laboratory equipment and foreign manufacturing experience exclusively produced 3510 atomic absorption spectrophotometer system, for system Xiangjie to quickly check the latest products, quotes, pictures, technical parameters, or product performance comparison to use very convenient.
3510 atomic absorption spectrophotometer system, u host economy
u high sensitivity and high stability of the graphite furnace system
u good optical system to ensure accurate and reliable measurements
u safe and reliable Flame Test
u High-Efficiency Analysis Workstation
Usage
3510 mainframe with built-in microcomputer systems, all working conditions set by the keyboard, instruments can automatically search for elements of the peak energy, Beijing time correction new cathode lamp and deuterium lamp energy self-balancing, and can handle data and maps, and truly reflects the atomic absorption analysis system for fast, accurate and sensitive characteristics of
Product parameter
Wavelength range: 190.0nm-860.0nm
Wavelength Accuracy: full-band ≤ ± 0.5nm
Wavelength repeatability: ≤ 0.3nm
Blazed wavelength: 250nm
Spectral Bandwidth: 0.1nm.0.2nm, 0.7nm, 1.4nm (switch selection)
Resolution: spectral bandwidth of 0.2nm, manganese can be separated from two-lane, that is, 279.5nm and 279.8nm, and the two lines between the energy value of less than 40% of the peak and trough
Reading mode: Continuous, maintaining, peak height, peak area
Background subtraction: the background signal 1ABS more than 30 times when you have the ability to background deduction
Absorbance Range: 0.000A-2.000A
Stability: Under normal circumstances, the instruments and copper lamps warm half an hour later, without firing a state of the copper lines 324.7nm determine baseline stability. 30-minute drift ≤ 0.004 Absorbance
Determination of typical elements: Flame, 5mg / L of Cu absorption is greater than 0.800A
Mexican Furnace Sensitivity: Cu1 × 10-11g (characteristic amount)
Precision: (relative standard deviation): ≤ 1%; ≤ 3% (Cu graphite furnace testing)